Summary of the results of the virtual fluoroscopy model

Method Patient 1 Patient 2 Patient 3 Patient 4
Proposed method Error (mm) 0.87 ± 1.2 0.56 ± 0.42 − 0.35 ± 0.70 − 0.01 ± 0.22
Correlation 0.998 0.999 0.999 0.999
Jaccard Index 0.950 (0.907–0.978) 0.982 (0.975–0.990) 0.949 (0.872–0.986) 0.988 (0.954–1.000)
Template matching
(soft)
Error (mm) − 13 ± 9.0 55 ± 30 − 12 ± 11 6.0 ± 9.1
Correlation 0.910 − 0.800 0.703 0.790
Template matching
(soft + bone)
Error (mm) − 18 ± 19 − 6.1 ± 12 − 0.69 ± 13 − 13 ± 14
Correlation 0.090 − 0.969 − 0.078 − 0.339

Error (mm) was converted according to the relationship 1.5 mm = 1 pixel