. 2016 Jun 24;17(1):444. doi: 10.4102/sajhivmed.v17i1.444

TABLE 4.

Performance of PIMA™ analysers compared to National Health Laboratory Services at different CD4+ T-cell thresholds.

CD4+ T-cells/mm3 Sensitivity (%) Specificity (%) Number misclassified Number misclassified (%) Correctly classified (%) FP Rate FN Rate Negative predictive value (%) Positive predictive value (%)
≤ 200 73.50 98.40 25 9.3 90.7 3/25 22/25 85.20 95.30
≤ 350 83.50 92.30 35 13.0 87.0 8/35 27/35 78.10 94.50
≤ 500 94.00 78.40 24 9.0 91.0 11/24 13/24 75.50 94.90

Note: PIMA™ point-of-care [POC] testing for CD4 counts in predicting antiretroviral initiation in HIV-infected individuals in KwaZulu-Natal, Durban, South Africa.

FP, false-positive; FN, false-negative.