Figure 6.

Figure 6

Dynamic range in dB for children who received their first cochlear implant at or before 12 months of age (early-implanted, depicted by filled circles) and children who received their first cochlear implant after 12 months of age (later-implanted, depicted by open squares). Data are shown for device settings at 6 months post-activation (6CI), and at chronological ages of 3 (Y3) and 5 years (Y5). DRs are denoted in charge in dB units (re:1nC) on the left y-axis and in clinical units CU (re:CIC4 implant and 25 μs pulse width) on the right y-axis of the graph. Vertical bars denote 95% confidence intervals.