Table 1.
Conversion of Clinical units (CU) to Charge in dB (re:1nC). For a current level of 200 clinical units (CU) and a pulse width of 25 μs, the current in microAmps (μA), charge in nanoCoulombs (nC) in dB re 1nC varied between implant types.
Example 1: CI24RE (CIC4 chip) |
Example 2: CI24M and CI24R (CIC3 chip) |
|||
---|---|---|---|---|
Current in microamps (μA) | μA = 17.5 * 100 (CU/255) | 648.14 | μA = 10 * 175(CU/255) | 574.44 |
Charge in nanocoulombs (nC) | (μA*10−6)* (μs*10−6)* 10−9 | 16.20 | (μA*10−6)* (μs*10−6)* 10−9 | 14.36 |
Charge in dB, re: 1nC | 20 × log (nC) | 24.19 | 20 × log (nC) | 23.14 |