. 2018;19(1):253–259. doi: 10.22034/APJCP.2018.19.1.253

Table 3.

Results for Right Lung Interface Tumor (RIT) Diameter of 2.5 cm

Evaluation Parameters Percentage Dose Difference w.r.t. Measured
3D-CRT IMRT-Static IMRT-Dynamic VMAT
AAA AcXB AAA AcXB AAA AcXB AAA AcXB
CAX Absolute Dose deviation (in %) 4 3 3 1.2 3.2 0.8 3.3 0.7
2D-Gamma (Pass%) at Isocentre 2%&2mm 92.3 94.1 91.7 95.2 91.9 95.58 90.1 94.1
3%&3mm 93.5 96.1 95.1 96.8 96.32 97.32 95.1 96.37
4%&4mm 95.2 96.93 96.1 97.8 97.35 98.14 96.3 98.04
Correlation Coefficient 0.9459 0.963 0.922 0.965 0.932 0.9689 0.946 0.988
2D Gamma (Pass% for 3%&3mm) at 4 cm from isocentre in normal lung 95.2 97 96.3 97.3 96.2 98.1 97.2 98.2