Skip to main content
. 2018 Feb 13;9(3):1130–1150. doi: 10.1364/BOE.9.001130

Table 5. Tolerance of HbT images recovered using two-composition-prior guided reconstruction to various experimental errors in all phantom cases.

(a) Metric 1: Overall RMSE ≤ 2 µM
Case 1 Case 2 Case 3 Case 4 Case 5 Mean std
Offset in x-direction (mm) 0.26 0.28 0.77 0.84 0.99 0.63 0.34
Source plate rotation (°) 0.12 0.14 0.33 0.33 0.41 0.27 0.13
Source plate L-R tilt (–°) −0.52 −0.49 −1.14 −1.13 −1.25 -0.91 0.37
Source plate L-R tilt ( + °) 0.58 0.36 1.00 1.17 1.41 0.90 0.43
Source plate N-C tilt (°) 0.13 0.08 0.34 0.42 0.63 0.32 0.22

Breast expansion (%) 0.55 0.57 1.45 1.37 1.68 1.12 0.53
Thickness increase (%) 1.22 0.78 1.43 1.51 1.98 1.38 0.44

Cross-talk (‰) 0.12 0.04 0.15 0.16 0.25 0.14 0.08
Light piping (‰) 0.52 0.35 0.92 1.07 1.84 0.94 0.58

(b) Metric 2: R, HbTt, or HbTn change ≤ 5%
Case 1 Case 2 Case 3 Case 4 Case 5 Mean std

Offset in x-direction (mm) 0.90 1.75 1.27 1.76 1.22 1.38 0.37
Source plate rotation (°) 0.42 0.79 1.56 1.56 1.34 1.13 0.51
Source plate L-R tilt (–°) −0.77 −0.66 −0.65 −0.65 −0.64 -0.67 0.05
Source plate L-R tilt ( + °) 0.70 0.69 0.67 0.69 0.67 0.68 0.01
Source plate N-C tilt (°) 0.64 0.39 0.37 0.39 0.44 0.45 0.11

Breast expansion (%) 3.91 4.16 1.91 >10.00* >10.00* 3.33** 1.23**
Thickness increase (%) 3.56 1.72 1.36 1.33 1.36 1.87 0.96

Cross-talk (‰) 0.70 0.42 0.23 0.17 0.36 0.38 0.21
Light piping (‰) 0.98 1.12 0.98 0.80 1.56 1.09 0.29