. 2018 Feb 13;9(3):1130–1150. doi: 10.1364/BOE.9.001130

Table 6. Tolerance of HbT images recovered without using structural priors to various experimental errors in all phantom cases.

(a) Metric 1: Overall RMSE ≤ 2 µM
Case 1 Case 2 Case 3 Case 4 Case 5 Mean std
Offset in x-direction (mm) 0.25 0.01 0.73 0.71 0.98 0.54 0.39
Source plate rotation (°) 0.11 0.01 0.31 0.27 0.40 0.22 0.16
Source plate L-R tilt (–°) −0.49 −0.02 −1.16 −0.97 −1.30 -0.79 0.53
Source plate L-R tilt ( + °) 0.56 0.01 1.02 1.02 1.55 0.83 0.58
Source plate N-C tilt (°) 0.11 0.00 0.29 0.28 0.60 0.26 0.23

Breast expansion (%) 0.50 0.02 1.30 1.13 1.66 0.92 0.66
Thickness increase (%) 1.09 0.04 1.28 1.21 1.87 1.10 0.66

Cross-talk (‰) 0.10 0.00 0.13 0.13 0.24 0.12 0.09
Light piping (‰) 0.41 0.11 0.80 0.84 1.71 0.77 0.60

(b) Metric 2: R, HbTt, or HbTn change ≤ 5%
Case 1 Case 2 Case 3 Case 4 Case 5 Mean std

Offset in x-direction (mm) 0.77 1.81 1.14 1.79 1.28 1.36 0.44
Source plate rotation (°) 0.38 0.79 1.47 1.72 1.38 1.15 0.55
Source plate L-R tilt (–°) −0.80 −0.60 −0.57 −0.64 −0.61 -0.64 0.09
Source plate L-R tilt ( + °) 0.88 0.63 0.59 0.68 0.63 0.68 0.12
Source plate N-C tilt (°) 0.51 0.41 0.40 0.42 0.46 0.44 0.05

Breast expansion (%) 4.97 4.19 >10.00* >10.00* >10.00* 4.58** 0.55**
Thickness increase (%) 3.36 1.76 1.49 1.39 1.41 1.88 0.84

Cross-talk (‰) 0.32 0.72 0.29 0.19 0.44 0.39 0.20
Light piping (‰) 0.96 1.94 1.16 0.85 1.77 1.34 0.49
*:

Tolerance values are beyond the maximal tested experimental error levels.

**:

Statistics are determined by excluding tolerance values marked by *.