. 2017 Sep 30;34(3):353–365. doi: 10.1007/s10554-017-1251-7

Table 5.

Optical coherence tomography analysis at 12 months

Stent type BiOSS LIM® Prolim® Alex®
Parameter n = 11 (%) n = 29 (%) n = 16 (%)
Stent apposition
 Embedded 95.8 98.6 99.2 0.00001*,***
 Protruding 2.5 1.2 0.8 0.0054*,***
 Uncovered 1.3 0.07 0.02 0.002*,***
 Malapposed 0.4 0.1 0 0.45
OCT parameters
 Mean minimal lumen area (mm2) 3.72 ± 0.57 4.82 ± 1.41 5.22 ± 1.95 0.027***
 Mean lumen area (mm2) 5.76 ± 0.73 6.21 ± 1.10 6.46 ± 1.8 0.35
 Mean stent area (mm2) 8.31 ± 0.4 8.39 ± 2.26 7.94 ± 1.84 0.7
 Mean neointima area (mm2) 2.55 ± 0.41 2.17 ± 0.37 1.49 ± 0.60 0.012*,***
 Neointima volume (mm3) 44.9 ± 4.9 28.16 ± 15.10 24.51 ± 17.64 0.0008*,***
 Mean neointima burden (%) 30.9 ± 6.2 24.6 ± 8.6 19.27 ± 8.11 0.0009***
Neoatherosclerosis assessment
 Type I (thin cap, lipid-rich) 1 (9.1) 1 (3.4) 0 0.92
 Type II (thick cap, layered) 2 (18.2) 4 (13.8) 2 (12.5) 0.97
 Type III (peristrut, homogenous) 6 (54.5) 10 (34.5) 3 (18.8) 0.25
 Type IV (preexisting, homogenous) 2 (18.2) 14 (48.3) 11 (68.7) 0.09

*p < 0.05 for Prolim vs. BiOSS LIM; **p < 0.05 Prolim vs. Alex; ***p < 0.05 BiOSS LIM vs. Alex