Table 1.
Designator | Test | Performance | |
---|---|---|---|
Tolerance | Action | ||
Quarterly | |||
QTPS1 | CPU/server | Functional | |
QTPS2 | Digitizer (if it is used clinically) | 2 mm | 3 mm |
QTPS3 | Electronic plan transfer | Data integrity | |
QTPS4 | Plan details | Data integrity | |
QTPS5 | Plotter/printer | 2 mm | 3 mm |
QTPS6 | Backup recovery | Functional | |
QTPS7 | CT geometry/density | 2 mm/0.02 | 3 mm/0.03 |