Table 3.

Monthly quality control tests

Designator Test Performance
Tolerance Action
Monthly
MK1 Mechanical stability and safety Functional
MK2 Cone/filters integrity and cone indicators Functional
MK3 Physical distance indicators 2 mm 3 mm
MK4 Accuracy of head tilt and rotation readouts 1.5°
MK5 Light/x ray field coincidence 2 mm 3 mm
MK6 Light field size 2 mm 3 mm
MK7 X ray field size indicator 2 mm 3 mm
MK8 X ray field uniformity 5% 8%
MK9 Output verification and reproducibility with head tilt and rotation 2% 4%
MK10 Timer accuracy verification 1% 2%
MK11 Dose rate output constancy 2% 4%
MK12 Records Complete