Table 4.
Designator | Test | Performance | |
---|---|---|---|
Tolerance | Action | ||
Annual | |||
AK1 | Reference dosimetry | 1% | 2% |
AK2 | Timer and end‐effect error | Characterize | ±0.05 min |
AK3 | Output linearity | n/a | 1% |
AK4 | Output reproducibility | 2% | 3% |
AK5 | Output error associated with beam interrupt | 2% | 4% |
AK6 | Beam quality | 10% | 15% |
AK7 | Alignment of focal spots | 0.5 mm | 1 mm |
AK8 | Focal spot size | Reproducible | |
AK9 | Percentage depth dose verification | Characterize and document | |
AK10 | Profiles verification | Characterize and document | |
AK11 | Independent quality control review | Complete |