Annual quality control tests

Designator Test Performance
Tolerance Action
Annual
AK1 Reference dosimetry 1% 2%
AK2 Timer and end‐effect error Characterize ±0.05 min
AK3 Output linearity n/a 1%
AK4 Output reproducibility 2% 3%
AK5 Output error associated with beam interrupt 2% 4%
AK6 Beam quality 10% 15%
AK7 Alignment of focal spots 0.5 mm 1 mm
AK8 Focal spot size Reproducible
AK9 Percentage depth dose verification Characterize and document
AK10 Profiles verification Characterize and document
AK11 Independent quality control review Complete