Notes on monthly tests

MK1 Verification that the unit and accessories are firmly anchored and may be used without endangering patients or staff
MK2 Verification of the integrity of the filters and cones and cone indicators
MK3 Verification of the optical and/or mechanical distance indicator if the unit is equipped with one
MK4 Verification of the angle readouts
MK5 Performance parameters refer to agreement at each edge. This test does not apply to all machine designs
MK6 Geometric test to verify the light field sizes (where applicable)
MK7 Confirmation of radiation field size when a variable collimation system is provided. At least two field sizes must be checked
MK8 Using a film, the flatness and symmetry of the X ray beam must be assessed for the largest cone
MK9 Quantitative dosimetric test: output reproducibility test at all energies with varying head tilt and rotation27
MK10 If the unit is equipped with a timer, its accuracy must be checked against a stop watch over a range of doses of 10−1000 cGy
MK11 Should be checked for all beam qualities for MU‐based systems
MK12 Documentation relating to the daily quality control checks, preventive maintenance, service calls and subsequent checks must be complete, legible, and the operator identified