MK1 |
Verification that the unit and accessories are firmly anchored and may be used without endangering patients or staff |
MK2 |
Verification of the integrity of the filters and cones and cone indicators |
MK3 |
Verification of the optical and/or mechanical distance indicator if the unit is equipped with one |
MK4 |
Verification of the angle readouts |
MK5 |
Performance parameters refer to agreement at each edge. This test does not apply to all machine designs |
MK6 |
Geometric test to verify the light field sizes (where applicable) |
MK7 |
Confirmation of radiation field size when a variable collimation system is provided. At least two field sizes must be checked |
MK8 |
Using a film, the flatness and symmetry of the X ray beam must be assessed for the largest cone |
MK9 |
Quantitative dosimetric test: output reproducibility test at all energies with varying head tilt and rotation27
|
MK10 |
If the unit is equipped with a timer, its accuracy must be checked against a stop watch over a range of doses of 10−1000 cGy |
MK11 |
Should be checked for all beam qualities for MU‐based systems |
MK12 |
Documentation relating to the daily quality control checks, preventive maintenance, service calls and subsequent checks must be complete, legible, and the operator identified |