. 2018 Jan 16;19(2):298–304. doi: 10.1002/acm2.12255

Table 1.

Imaging parameters for stability and sensitivity evaluation of the contrast response measurement. Baseline parameters list the nominal, starting point scan parameters. Stability and sensitivity parameters match the baseline parameters except for the indicated changes

Contrast response test 9L C1‐5 S1‐5
Baseline
Transmit frequency 7 MHz 4 MHz 3 MHz
Power 60% 60% 60%
Gain 85 dB 65 dB 60 dB
Dynamic range 69 dB 84 dB 84 dB
Gray map D D D
Phantom background 0.7 dB/cm/MHz gel 0.7 dB/cm/MHz gel 0.7 dB/cm/MHz gel
Stability
Changed gain/power 80/60%, 85/100%, 85/30%, 90/60% 60/60%, 65/100%, 65/30%, 70/60% 55/60%, 60/100%, 60/30%, 65/60%
Changed gray map A and J A and J A and J
Changed frequency 9 MHz 5 MHz 4 MHz
Sensitivity
Changed dynamic range 48 and 90 dB 75 and 96 dB 76 and 96 dB