Skip to main content
. 2018 Jan 16;19(2):298–304. doi: 10.1002/acm2.12255

Table 2.

Imaging parameters for stability and sensitivity evaluation of the depth of penetration (DOP) measurement. Baseline parameters list the starting point scan parameters. Stability and sensitivity parameters match the baseline parameters except for the indicated changes. The second sensitivity test used the baseline scan parameters but with the 0.5 dB/cm/MHz gel region of the CIRS phantom

Depth of penetration test 11L C1‐5 S1‐5
Baseline
Transmit frequency 11 MHz 5 MHz 4 MHz
Power 100% 100% 100%
Gain 80 dB 85 dB 75 dB
Dynamic range 90 dB 90 dB 90 dB
Gray map D D D
Phantom background 0.7 dB/cm/MHz gel 0.7 dB/cm/MHz gel 0.7 dB/cm/MHz gel
Stability
Changed gain/dynamic range 75/90, 80/84, 80/96, 85/90 80/90, 85/84, 85/96, 90/90 70/90, 75/84, 75/96, 80/90
Changed gray map A and J A and J A and J
Sensitivity
Changed power 50% and 20% 50% and 20% 50% and 20%
Changed phantom background 0.5 dB/cm/MHz gel 0.5 dB/cm/MHz gel 0.5 dB/cm/MHz gel