Spatial resolution test 9L C1‐5 S1‐5
Baseline
Transmit frequency 7 MHz 4 MHz 3 MHz
Power 100% 100% 100%
Gain 85 dB 63 dB 60 dB
Dynamic range 90 dB 90 dB 90 dB
Gray map D D D
Phantom background 0.7 dB/cm/MHz gel 0.7 dB/cm/MHz gel 0.7 dB/cm/MHz gel
Stability
Changed gain/dynamic range 80/90, 85/84, 85/96, 90/90 58/90, 63/84, 63/96, 68/90 55/90, 60/84, 60/96, 65/90
Changed gray map A and J A and J A and J
Sensitivity
Changed transmit frequency 5 and 9 MHz 2 and 5 MHz 2 and 4 MHz