Table 1.

Limits of Detection Testing of the RealTime Hepatitis C Virus (HCV) Viral Load Assay Using RNA Extracted From Circular Punches of Mock Dried Blood Spots Made With Whole Blood Containing Various Target Concentrations of HCV RNA

Target Concentration, IU/mL Whole Blood 6-mm DBS Punch 12-mm DBS Punch
Tested, No. Detected, No. % Tested, No. Detected, No. %
5000 15 15 100 31 31 100
2000 19 19 100 16 16 100
1000 30 27 90 40 40 100
500 30 21 70 33 31 94
250 28 9 32 28 21 75

Abbreviations: DBS, dried blood spot; IU, international units.