Table 2.

Bland and Altman analysis of wireless heart rate (HR) and respiratory rate (RR) versus reference monitor in postoperative patients

Parameter No of measurement pairs Bias SD Lower 95% LoA Upper 95% LoA
Complete dataset
 HR 55 565 −1.1 3.8 −8.8 6.5
 RR 56 674 −2.3 6.8 −15.8 11.2
Filtered dataset*
 HR 3986 −1.2 2.2 −5.7 3.2
 RR 4001 −2.4 4.2 −10.8 5.9

*Dataset after applying a median filter.

LoA, limits of agreement.