Skip to main content
. 2018 Feb 24;23(2):126–135. doi: 10.1016/j.rpor.2018.01.005

Table 5.

Dosimetric results of treatment plans computed on pCT and sCBCT image sets using Octavius 1500 ion chamber array.

Image Sets Central axis dose difference (%)
γ-Evaluation pass rate (%)
IMRT VMAT IMRT
VMAT
3% DD/3 mm DTA
2% DD/2 mm DTA
3% DD/3 mm DTA
2% DD/2 mm DTA
Local Global Local Global Local Global Local Global
pCT −1.12 0.27 88.6 96.5 69.6 87.1 95.1 99.2 80.3 93.4
S1-CBCT −1.01 0.43 89.5 96.8 71.6 88.2 93.8 99.2 78.8 93.8
S2-CBCT −1.16 0.60 89.9 96.9 71.7 88.2 94.5 99.2 79.4 93.9