Table 3.

Annual quality control tests

Designator Test Performance
Action
Annually
A1 Hand crank operation Functional
A2 Leakage radiation Reproducible
A3 Multi‐channel indexer function Functional
A4 Dwell time accuracy 1%
A5 Timer linearity 1%
A6 Transit time/transit dose reproducibility Reproducible
A7 Dosimetric length of applicators and guide tubes 1 mm
A8 Applicators and templates dimensions Reproducible
A9 Shield integrity of shielded applicators Reproducible
A10 X ray marker positional accuracy 1 mm
A11 Document staff review of emergency response procedures Complete
A12 Independent quality control review Complete
A13 PDR sequencing (for PDR only) Functional