Table 3.
Designator | Test | Performance |
---|---|---|
Action | ||
Annually | ||
A1 | Hand crank operation | Functional |
A2 | Leakage radiation | Reproducible |
A3 | Multi‐channel indexer function | Functional |
A4 | Dwell time accuracy | 1% |
A5 | Timer linearity | 1% |
A6 | Transit time/transit dose reproducibility | Reproducible |
A7 | Dosimetric length of applicators and guide tubes | 1 mm |
A8 | Applicators and templates dimensions | Reproducible |
A9 | Shield integrity of shielded applicators | Reproducible |
A10 | X ray marker positional accuracy | 1 mm |
A11 | Document staff review of emergency response procedures | Complete |
A12 | Independent quality control review | Complete |
A13 | PDR sequencing (for PDR only) | Functional |