TABLE 2.

Size of puncture and penumbra of tissue blanching measured under microscopy comparing radiofrequency-assisted needle and guidewire punctures in different cardiac tissue in freshly explanted pig hearts

Tissue Electrified traversals Mechanical traversals
Needle electrical, n = 18 Astato electrical (tip), n = 18 Needle mechanical, n = 9 Astato mechanical (back-end), n = 9
Atrial septum Hole (mm) 1.11 ± 0.40 0.37 ± 0.08 P = .009 0.63 ± 0.11 0.19 ± 0.16 P = .01
Penumbra (mm) 3.62 ± 1.23 0.72 ± 0.29 P = .003 Negligible Negligible
Subgroup
Needle 50 W, n = 9 Needle 15W, n = 9 Astato 50 W, n = 9 Astato 15 W, n = 9
Hole (mm) 0.99 ± 0.38 1.24 ± 0.46 P = .65 0.39 ± 0.11 0.36 ± 0.05 P = .80
Penumbra (mm) 3.68 ± 0.74 3.57 ± 1.80 P = .89 0.73 ± 0.22 0.70 ± 0.40 P = .93
Ventricular septum Hole (mm) 1.24 ± 0.24 0.45 ± 0.05 P < .001 1.07 ± 0.24 0.13 ± 0.23 P = .01
Penumbra (mm) 5.30 ± 1.16 0.75 ± 0.20 P < .001 Negligible Negligible
Subgroup
Needle 50 W, n = 9 Needle 15W, n = 9 Astato 50 W, n = 9 Astato 15 W, n = 9
Hole (mm) 1.11 ± 0.24 1.37 ± 0.20 P = .08 0.47 ± 0.02 0.42 ± 0.06 P = .38
Penumbra (mm) 5.82 ± 0.44 4.77 ± 1.52 P = .44 0.86 ± 0.15 0.63 ± 0.19 P = .14
Aorta Hole (mm) 0.88 ± 0.16 0.29 ± 0.12 P < .001 0.70 ± 0.11 0.12 ± 0.12 P = .007
Penumbra (mm) 2.77 ± 1.20 0.59 ± 0.15 P = .005 Negligible Negligible
Subgroup
Needle 50 W, n = 9 Needle 15W, n = 9 Astato 50 W, n = 9 Astato 15 W, n = 9
Hole (mm) 0.89 ± 0.15 0.87 ± 0.15 P = .92 0.33 ± 0.12 0.25 ± 0.12 P = .35
Penumbra (mm) 3.12 ± 1.67 2.41 ± 0.68 P = .38 0.58 ± 0.17 0.60 ± 0.17 P = .04

A P value of <.05 was considered statistically significant.