TABLE I.

Main characteristics of the four devices compared in this study. SPL represents the sound pressure level, f0 represents the fundamental frequency, and Dt represents the phonation time percentage.

Device Type of transducer Frame length Calibration procedure for SPL Estimated parameters Algorithm for f0 evaluation Uncertainty reported in literature
VocaLog2 Contact microphone 1 s Yes SPL NA SPL mean error of 1.3–1.9 dB in a vowel task and of 1.5–2.4 dB in a reading task
VoxLog Accelerometer and calibrated air microphone 0.1 s, 1 s, 5 s, 30 s, 60 s, 180 s and 300 s No A-weighted SPL for voice and noise, f0, Dt Fast Fourier Transform- based SPL mean error 4.3 dB, f0 mean error of −0.56 Hz
Voice Care Contact microphone 0.03 s Yes SPL, f0, Dt Autocorrelation-based SPL mean error −1.6 dB and standard deviation 2.5 dB, f0 absolute error ≤ 3 Hz
APM3200 Accelerometer 0.05 s Yes SPL, f0, Dt and other parameters derived from the previous ones Autocorrelation-based SPL mean error 3.2 dB with a standard deviation of 6 dB, f0 error 1 Hz