Table 1.

ICD Programming Schema

Manufacturer VT Zone Detection VT Zone Therapy VF Zone Detection VF Zone Therapy
Medtronic Inc Rate: 180 bpm
100 intervals (33 s) to detection
ATP×5, 25 J×2 Rate: 222 bpm
120/160 intervals to detection (32.4 s)
25 J, 35 J×5
Boston Scientific Inc Rate: 180 bpm
30 s to detection
ATP×8, 21 J, 41 J×6 Rate: 220 bpm
15 s to detection
29 J, 41 J×7
St. Jude Medical Rate: 180 bpm
100 intervals (33 s) to detection
ATP×3, 36 J, 40 J×2 Rate: 240 bpm
100 intervals to detection (25 s)
36 J, 40 J×5

ICD programming schema for the three device manufacturers involved in the trial. A number followed by “J” indicates a shock therapy with the number referring to the shock energy in joules (J). ATP indicates anti‐tachycardia pacing; bpm, beats per minute; ICD, implantable cardioverter‐defibrillator; VF, ventricular fibrillation; VT, ventricular tachycardia.