Table 5.
Condition | No. of paired CGM system-YSI reference readings | Percent within 15/15% reference | Percent within 20/20% reference | Percent within 30/30% reference | Percent within 40/40% reference | MARD (%) [95% CI] |
---|---|---|---|---|---|---|
Exercise condition | 1190 | 87.6 | 95.1 | 99.7 | 99.9 | 8.3 (7.5–9.3) |
Nonexercise condition | 15,463 | 85.6 | 93.2 | 97.9 | 99.3 | 8.8 (8.1–9.3) |
In-clinic accuracy is assessed compared with venous YSI reference measurement. The effect of exercise on accuracy was evaluated by instructing participants with a single sensor to perform upper arm exercises for 30 min. Results compared with a nonexercise condition are reported. The percentage of system readings within ±15 mg/dL or 15% of YSI reference values (15/15%), ±20 mg/dL or 20% of YSI reference values (20/20%), ±30 mg/dL or 30% of YSI reference values (30/30%), or ±40 mg/dL or 40% of YSI reference values (40/40%) are reported.