Table 6.

Effect of System Rate of Change on Continuous Glucose Monitoring System and Reference Agreement

CGM ROC [mg/(dL·min)] No. of paired CGM system-YSI reference readings Percent within 15/15% reference Percent within 20/20% reference Percent within 30/30% reference Percent within 40/40% reference MARD (%) [95% CI]
Less than −2 251 80.9 90.8 95.6 98.4 10.5 (8.2–14.1)
−2 to −1 1285 82.8 91.7 96.9 98.7 10.1 (8.9–11.4)
−1 to 1 12,740 86.8 94.1 98.4 99.5 8.5 (7.7–8.8)
1 to 2 1188 80.6 89.7 97.8 99.7 9.4 (8.8–10.3)
>2 515 81.9 91.1 97.5 99.4 9.4 (8.3–10.2)

In-clinic accuracy is assessed compared with venous YSI reference measurement. The percentage of system readings within ±15 mg/dL or 15% of YSI reference values (15/15%), ±20 mg/dL or 20% of YSI reference values (20/20%), ±30 mg/dL or 30% of YSI reference values (30/30%), or ±40 mg/dL or 40% of YSI reference values (40/40%) are reported.