Table 8.
In-Clinic Hypoglycemic and Hyperglycemic Event Detection Using Both Threshold and 10 Minutes Predictive Alerts
| Glucose setting (mg/dL) | Confirmed event detection rate | Missed event detection rate | True alert rate | False alert rate |
|---|---|---|---|---|
| Hypoglycemic alert | ||||
| 70 | 93% | 7% | 86% | 14% |
| Hyperglycemic alert | ||||
| 180 | 96% | 4% | 94% | 6% |