Skip to main content
. 2018 Mar 1;20(3):197–206. doi: 10.1089/dia.2017.0142

Table 8.

In-Clinic Hypoglycemic and Hyperglycemic Event Detection Using Both Threshold and 10 Minutes Predictive Alerts

Glucose setting (mg/dL) Confirmed event detection rate Missed event detection rate True alert rate False alert rate
Hypoglycemic alert
 70 93% 7% 86% 14%
Hyperglycemic alert
 180 96% 4% 94% 6%