. 2018 Mar 16;8(3):171. doi: 10.3390/nano8030171

Table 3.

Comparison of amperometric response of varieties of nanoporous gold (np-Au)-based glucose biosensors.

Electrode Rf Potential a Mediator Linear (mM) LOD (µM) Sensitivity
(µA·cm−2·mM−1)
Ref.
GOx-Chi/PB/np-Au/Ti NA −1.0 V b PB 0.1–2.0 2.5 177 c [119]
Naf/GOx/np-Au/GC NA 0.4 V - 1–18 196 0.049 c [30]
GOx/SAM/np-Au/GC NA −0.2 V b - 3–8 10 8.6 [110]
Naf/GOx/PB/np-Au/Cr/Si 40 0 V b PB 2–30 300 50 [120]
GOx/Os(bpy)2P/np-Au/SiO2 NA NA Os(bpy)2P NA 2 75 [121]
GOx/PEDOT/np-Au/GC NA 0.2 V BQ 0.1–15 10 7.3 [122]
GOx/np-Au/Au/Si 36 NA - 0.1–0.5 73 21.14 [123]
GOx/GA/CA/np-Au/GC 7 0.2 V
0.3 V
BQ
FCA
1–10
1–10
NA
NA
3.53
1.35
[124]
GOx/DTDPA/np-Au/GC 8 0.2 V BQ 1–10 NA 2.187 [9]
GOx/np-Au/GC 8 0.3 V - 0.05–10 1.02 12.1 [125]

a vs. SCE, b vs. Ag/AgCl, c µA·mM−1, Rf: roughness factor of np-Au, Chi: chitosan, GC: glassy carbon electrode, Naf: Nafion, PB: Prussian blue (ferric hexacyanoferrate), DET: direct electron transfer, BQ: p-benzoquinone, FCA: ferrocenecarboxylic acid, GA: glutaraldehyde, CA: cysteamine, DTDPA: 3,3′-Dithiodipropionic acid, PEDOT: poly(3,4-ethylenedioxythiophene), Os(bpy)2P: [Os(2,2′-bipyridine)2(polyvinylimidazole)Cl]+/2+.