Skip to main content
. 2018 Mar 26;56(4):e01523-17. doi: 10.1128/JCM.01523-17

TABLE 2.

Stability tests using panels stored at different temperatures for 3 months

Panel storage temp and strain MIC (μg/ml)
Expected range T0 T7 T14 T30 T90
4°C
    ATCC 25922 0.25–2a 1 0.5 0.25 0.5 0.5
    ATCC 27853 0.5–4a 1 1 1 1 1
    CSR-55 4–16 16 8 8 8 8
    CSR-57 1–4 4 4 2 4 4
    CSR-68 8–32 ≥16 ≥16 ≥16 ≥16 ≥16
    NCTC 13846 2–8a 4 16 4 4 8
    Negative control NGb NG NG NG NG
Room temp
    ATCC 25922 0.25–2a 0.5 0.5 0.5 1 1
    ATCC 27853 0.5–4a 1 1 1 1 1
    CSR-55 4–16 8 16 8 16 8
    CSR-57 1–4 4 4 2 4 4
    CSR-68 8–32 ≥16 ≥16 ≥16 ≥16 ≥16
    NCTC 13846 2–8a 4 4 4 4 4
    Negative control NG NG NG NG NG
35°C
    ATCC 25922 0.25–2a 0.5 0.5 0.5 1 1
    ATCC 27853 0.5–4a 1 1 1 1 2
    CSR-55 4–16 8 8 8 8 ≥16
    CSR-57 1–4 4 4 4 4 4
    CSR-68 8–32 ≥16 ≥16 ≥16 ≥16 ≥16
    NCTC 13846 2–8a 4 4 4 4 8
    Negative control NG NG NG NG NG
a

According to EUCAST Routine and Extended Internal Quality Control for MIC Determination and Disk Diffusion (version 7.0, valid from 1 January 2017).

b

NG, no growth.