. 2018 Mar 26;56(4):e01329-17. doi: 10.1128/JCM.01329-17

TABLE 2.

Monomicrobial call performance comparison with FDA-cleared Accelerate Pheno system software and postclearance 2017 software update

Parametera Value for software
FDA-cleared 2017 software update
No. of total valid samples 793 872
Total no. of MONO by Accelerate PhenoTest BC kit 557 775
Total no. of MONO by Accelerate PhenoTest BC kit and confirmed by reference 545 754
No. of false MONOs 12 21
    No. of false MONO, resolved by Gram staining 10 16
    No. of false MONO, unresolved by Gram staining 2 5
Positive predictive value (95% CI) 97.8 (96.3–98.8) 97.3 (95.9–98.2)
Positive predictive value after resolving with Gram staining (95% CI) 99.6 (98.7–99.9) 99.4 (98.5–99.7)
a

MONO, monomicrobial calls. The positive predictive value is the percentage correct.