Published in final edited form as: IEEE Trans Ultrason Ferroelectr Freq Control. 2017 Aug 14;64(10):1542–1557. doi: 10.1109/TUFFC.2017.2739649

TABLE I.

Values for comparison of output, including power supply voltage (VDC), measured peak-peak voltage on the transducer (Vpp), pulse-average acoustic power (Ppa), estimated surface pressure and intensity (p0 and I0), and peak focal pressure amplitudes (p+ and p−).

VDC(V) Vpp(V) Ppa(W) p0 (kPa) I0 (W/m2) p+ (kPa) p− (kPa)
2.5 6.3 0.72 12 48 763 718
5 12.6 2.8 24 192 1575 1397
10 25.3 11.2 48 767 3374 2650
50 126.3 271 238 19180 53003 96956
70 176.8 525 333 37600 76638 12149
75 189.4 593 358 43150 79705 12711
100 252.5 1063 477 76700 89813 15120
150 378.8 2346 715 172600 101289 18833
200 505 4400 954 306800 108161 21689