Table 1.

Knee MR acquisition parameters of sequences assessed for our study according to the OAI study protocol (adapted from Peterfy CG et al.[29]). Coil: quadrature transmit-receive coil (USA Instruments, Aurora, OH)

Scan COR IW 2D TSE SAG 3D DESS WE1 COR T1W 3D FLASH WE SAG IW 2D TSE FS2
Plane Coronal Sagittal Coronal Sagittal
Fat suppression No WE WE FS
Matrix (phase/frequency) 307/384 307/384 512/512 313/448
No. of slices 35 160 80 37
FOV (mm) 140 140 160 160
Slice thickness/ gap (mm) 3/0 0.7/0 1.5/0 3/0
Flip angle (°) 180 25 12 180
TE/TR (ms) 29/3700 4.7/16.3 7.57/20 30/3200
Bandwidth (Hz/pixel) 352 185 130 248
No. of excitations (averaged) 1 1 1 1
Echo-train length 7 1 1 5
Acquisition time (min) 3.4 10.6 8.6 4.7

COR = coronal; DESS = dual-echo in steady state; FLASH = fast low-angle shot; FS = fat suppression; TSE = turbo spin-echo; FOV = field of view; IW = intermediate-weighted; SAG = sagittal; TE = echo time; TR = repetition time; WE = water excitation.

1

also used for axial and coronal multi-planar reformations (MPR); slice thickness, 1.5mm each.

2

used for the assessment of SPFP imaging characteristics.