. 2016 Sep 8;17(5):509–522. doi: 10.1120/jacmp.v17i5.6287

Table 4.

Air‐kerma rate and barrier thickness percentage errors between RadShield and manual approach: catheterization lab.

Wall Number Kerma Rate Lead Concrete Gypsum Wood Steel Glass
1 0.67% 0.01% 2.83% 2.35% 1.18% 2.77% 1.25%
2 1.88% 0.50% 2.53% 2.03% 1.42% 2.35% 0.97%
3 3.58% 1.81% 0.52% 1.02% 4.52% 3.07% 0.19%
4 0.96% 1.27% 2.58% 2.10% 1.33% 2.62% 1.05%
5 0.90% 0.81% 2.33% 1.86% 1.59% 3.34% 1.08%
Average 1.60% 0.88% 2.16% 1.87% 2.01% 2.83% 0.91%