. 2016 Sep 8;17(5):184–199. doi: 10.1120/jacmp.v17i5.6260

Table 4.

Summary of dosimetric results in terms of beam delivery efficiency in CyberKnife InCise MLC‐based and cone/Iris‐based plans. Data represent mean±SD (median). The number of plans for each category are identical to Table 3. Wilcoxon signed‐rank p was below 0.05 for all cases.

MU Delivery Time (min) Number of Beams Ratio (MLC/cone or Iris)
Cone/Iris MLC Cone/Iris MLC Cone/Iris MLC beam MLC Segment MU Delivery Time (min)
All cases
21275±10985
8971±5489
58±18
37±12
182±85
63±34
105±64
0.42±0.16
0.65±0.12
(17041) (9091) (58) (36) (166) (58) (86) (0.40) (0.66)
    single target
18149±11091
6709±3496
51±14
32±9
168±94
50±25
88±63
0.40±0.18
0.65±0.14
(15178) (4752) (51) (31) (151) (36) (52) (0.36) (0.68)
    multiple targets
26832±8777
12993±6248
72±15
46±13
207±61
86±37
135±58
0.47±0.10
0.64±0.07
(26441) (11727) (67) (44) (202) (73) (138) (0.49) (0.65)
    circular shape
18355±8115
8325±4976
56±17
36±11
168±17
61±33
94±58
0.45±0.16
0.66±0.12
(16806) (6572) (56) (33) (56) (49) (86) (0.43) (0.66)
    irregular shape of target or abutting with OARs
28782±14309
10634±6772
65±19
40±17
218±112
68±38
132±76
0.37±0.15
0.61±0.11
(26442) (10902) (62) (40) (217) (65) (145) (0.30) (0.65)
     target volume1cc
14606±5617
6329±3651
47±15
31±8
108±49
43±17
53±27
0.42±0.10
0.68±0.13
(13280) (3973) (46) (29) (110) (37) (47) (0.40) (0.66)
     1cc<target volume5cc
22795±11122
9671±8267
64±21
38±18
177±65
66±49
95±67
0.37±0.14
0.58±0.11
(16666) (4845) (62) (32) (156) (35) (53) (0.30) (0.61)
     target volume>5cc
24941±12569
10326±4819
61±13
41±6
245±79
74±20
156±45
0.49±0.20
0.69±0.11
(21632) (10902) (62) (40) (242) (65) (145) (0.52) (0.69)

MU = monitor unit; OAR=organ at risk.