Skip to main content
. 2016 Sep 8;17(5):235–244. doi: 10.1120/jacmp.v17i5.6204

Table 2.

Average, maximum, minimum, and standard deviation (SD) of the static MLC deviation detection across five different leaf pairs as shown in Fig 7.

Introduced Deviation (mm) Measurement
Avg. (mm) Min. (mm) Max. (mm) SD (mm)
1.00 0.45 0.37 0.50 0.05
2.00 1.83 1.79 1.88 0.04
3.00 2.60 2.54 2.68 0.06
4.00 3.42 3.32 3.55 0.10
5.00 4.91 4.78 5.03 0.09