. 2016 Sep 8;17(5):20–33. doi: 10.1120/jacmp.v17i5.6118

Table 4.

The patient QA breakdown by technique, MLC type, LINAC model, device, and operator. Note that MLC type comparison is for the QAs delivered in the TrueBeam LINACs only (1,136 in total), LINAC model comparison excludes the plans calculated in the TPS with the Elekta Versa LINAC model (3,082 in total), and operator comparison excludes the operators who performed the lower number of QAs (2,796 in total).

Technique MLC‐type LINAC Model Device Operator
IMRT VMAT HD Non‐HD TrueBea m 2100 #1 #2 #1 #2
# of QAs 1601 1515 492 644 809 2273 1552 1564 1382 1414
# of QAs for Γ<90% 99 41 3 6 6 134 65 75 54 61
% failures 6.2% 2.7% 0.6% 0.9% 0.7% 5.9% 4.2% 4.8% 3.9% 4.3%
p‐value
<<0.001
<<0.001
<<0.001
0.39 0.64

HD = high‐definition; 2100 = Varian Clinac 2100; Γ=gamma passing rate (3%/3 mm).