. 2016 Sep 8;17(5):466–481. doi: 10.1120/jacmp.v17i5.6262

Table 1.

Summary of dose uncertainties in percentage (%).

Characteristic Red Channel Green Channel Blue Channel
Response curves and fitting procedure 2.6 4.3 4.1
Dose resolution of the system 1.8 2.3 3.1
Film reproducibility 0.2 0.3 0.3
Film uniformity 0.2 0.3 0.3
Relative orientation of the film 6.2 2.7 3.3
Reproducibility of the response of the scanner 0.3 0.3 0.3
Homogeneity on the bed of scanner 2.0 3.0 4.5
Total Uncertainty a 3.2 4.9 5.2
a

Without considering relative orientation of the film and homogeneity on the bed of the scanner.