Skip to main content
. 2016 Sep 8;17(5):245–261. doi: 10.1120/jacmp.v17i5.6282

Table 5.

A subset of the OCFs covering the range of energies, field sizes, and applicators measured at 100 cm SSD. “N/A” depicts physically impossible setups.

OCF 6 MeV 12 MeV 20 MeV
Field Size (cm2) Applicator (cm2) Applicator (cm2) Applicator (cm2)
6×6
15×15
25×25
6×6
15×15
25×25
6×6
15×15
25×25
2×2
1.000 0.993 0.992 1.005 0.998 1.000 1.003 0.991 0.991
3×3
1.007 1.002 0.995 1.005 1.000 0.988 1.007 0.995 0.986
4×4
1.004 1.008 0.999 1.004 1.003 0.994 1.007 0.997 0.988
6×6
N/A 1.002 0.997 N/A 1.001 0.995 N/A 1.003 0.994
8×8
N/A 0.999 0.997 N/A 1.000 0.995 N/A 1.003 0.994
10×10
N/A 0.998 0.998 N/A 0.999 0.995 N/A 1.001 0.996
12×12
N/A 0.996 0.997 N/A 0.996 0.995 N/A 1.000 0.997
15×15
N/A N/A 0.995 N/A N/A 0.994 N/A N/A 0.995
20×20
N/A N/A 0.996 N/A N/A 0.995 N/A N/A 0.995