Table 5.
A subset of the OCFs covering the range of energies, field sizes, and applicators measured at 100 cm SSD. “N/A” depicts physically impossible setups.
| OCF | 6 MeV | 12 MeV | 20 MeV | ||||||||||||||
|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
| Field Size (cm2) | Applicator (cm2) | Applicator (cm2) | Applicator (cm2) | ||||||||||||||
|
|
|
|
|
|
|
|
|
|
|||||||||
|
|
1.000 | 0.993 | 0.992 | 1.005 | 0.998 | 1.000 | 1.003 | 0.991 | 0.991 | ||||||||
|
|
1.007 | 1.002 | 0.995 | 1.005 | 1.000 | 0.988 | 1.007 | 0.995 | 0.986 | ||||||||
|
|
1.004 | 1.008 | 0.999 | 1.004 | 1.003 | 0.994 | 1.007 | 0.997 | 0.988 | ||||||||
|
|
N/A | 1.002 | 0.997 | N/A | 1.001 | 0.995 | N/A | 1.003 | 0.994 | ||||||||
|
|
N/A | 0.999 | 0.997 | N/A | 1.000 | 0.995 | N/A | 1.003 | 0.994 | ||||||||
|
|
N/A | 0.998 | 0.998 | N/A | 0.999 | 0.995 | N/A | 1.001 | 0.996 | ||||||||
|
|
N/A | 0.996 | 0.997 | N/A | 0.996 | 0.995 | N/A | 1.000 | 0.997 | ||||||||
|
|
N/A | N/A | 0.995 | N/A | N/A | 0.994 | N/A | N/A | 0.995 | ||||||||
|
|
N/A | N/A | 0.996 | N/A | N/A | 0.995 | N/A | N/A | 0.995 | ||||||||