Skip to main content
. 2016 Sep 8;17(5):245–261. doi: 10.1120/jacmp.v17i5.6282

Table 7.

A subset of the OCFs covering the range of energies, field sizes, and applicators measured at 110 cm SSD. “N/A” depicts physically impossible setups, whereas dashes denote physically available combinations that were not measured.

OCF 6 MeV 12 MeV 20 MeV
Field Size (cm2) Applicator (cm2) Applicator (cm2) Applicator (cm2)
6×6
15×15
25×25
6×6
15×15
25×25
6×6
15×15
25×25
2×2
1.005 0.993 0.990 0.994 0.996 0.997 1.002 0.997 0.993
4×4
1.004 1.006 1.003 1.004 1.005 1.002 1.001 1.000 0.990
6×6
N/A 1.001 N/A 0.998 N/A 0.992
8×8
N/A 1.006 1.004 N/A 1.005 1.002 N/A 1.002 0.995
10×10
N/A 1.004 N/A 1.002 N/A 0.997
12×12
N/A 1.000 1.002 N/A 0.999 1.000 N/A 1.000 0.997
15×15
N/A N/A 1.000 N/A N/A 0.999 N/A N/A 0.999
20×20
N/A N/A 1.000 N/A N/A 1.000 N/A N/A 1.000