Table 7.
A subset of the OCFs covering the range of energies, field sizes, and applicators measured at 110 cm SSD. “N/A” depicts physically impossible setups, whereas dashes denote physically available combinations that were not measured.
| OCF | 6 MeV | 12 MeV | 20 MeV | ||||||||||||||
|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
| Field Size (cm2) | Applicator (cm2) | Applicator (cm2) | Applicator (cm2) | ||||||||||||||
|
|
|
|
|
|
|
|
|
|
|||||||||
|
|
1.005 | 0.993 | 0.990 | 0.994 | 0.996 | 0.997 | 1.002 | 0.997 | 0.993 | ||||||||
|
|
1.004 | 1.006 | 1.003 | 1.004 | 1.005 | 1.002 | 1.001 | 1.000 | 0.990 | ||||||||
|
|
N/A | ‐ | 1.001 | N/A | ‐ | 0.998 | N/A | ‐ | 0.992 | ||||||||
|
|
N/A | 1.006 | 1.004 | N/A | 1.005 | 1.002 | N/A | 1.002 | 0.995 | ||||||||
|
|
N/A | ‐ | 1.004 | N/A | ‐ | 1.002 | N/A | ‐ | 0.997 | ||||||||
|
|
N/A | 1.000 | 1.002 | N/A | 0.999 | 1.000 | N/A | 1.000 | 0.997 | ||||||||
|
|
N/A | N/A | 1.000 | N/A | N/A | 0.999 | N/A | N/A | 0.999 | ||||||||
|
|
N/A | N/A | 1.000 | N/A | N/A | 1.000 | N/A | N/A | 1.000 | ||||||||