Skip to main content
. 2016 Sep 8;17(5):366–376. doi: 10.1120/jacmp.v17i5.6327

Table 3.

REM500 measurements of initial spot scanning plans.

Field Energy (MeV) Range (cm) SOBP (cm) Gantry Angle Couch Angle Plan REM500 (μSv) Statistical Error
RPO 1144.9‐72.5 14.8 13.3 260 0 Plan 1 1.5 13%
LAO 146.9‐76.1 14.7 12.78 50 0 Plan 1 1.8 13%
LPO 1138.1‐72.5 13.9 12.84 105 0 Plan 1 0.8 17%
RPO 1148.8‐74.3 15.8 12.34 255 0 Plan 2 1.9 10%
RAO 1139.8‐72.5 14.6 13.12 300 0 Plan 2 1.1 15%
Vertex 1178.6‐119.6 21.3 12.09 290 90 Plan 2 41.1 2%

SOBP=spreadout Bragg peak.