Table 1.
Minimum SRS‐SBRT relevant equipment QA and tolerances for C‐arm linac systems
Frequency | Test | Tolerance |
---|---|---|
Daily | Laser localization — only if using SRS techniques relying on lasers for target localization (e.g., frame‐based SRS without X‐ray IGRT) | 1 mm |
Collimator size indicator for clinically relevant aperture | 2 mm total | |
Radiation isocentricity test (limited gantry and couch positions) — maximum deviation in center of target object relative to each projection's beam central axis | 1.0 mm SRS, 1.5 mm SBRT | |
IGRT positioning/repositioning | 1 mm SRS, 2 mm SBRT | |
Imaging subsystem interlocks | Functional | |
Stereotactic interlocks — cone size, backup jaws | Functional | |
Accelerator output constancy | ±3% | |
Monthly |
Radiation isocentricity test — covering complete range of gantry, couch, collimator positions used clinically — maximum deviation in center of target object relative to each projection's beam central axis
*Note: If both MLC and fixed conical collimators are used, both must be evaluated at least monthly |
1.0 mm SRS, 1.5 mm SBRT |
Treatment couch position indicators: relative over the maximum clinical range | 1 mm/0.5° | |
Output constancy at relevant dose rates | 2% | |
Annually | SRS arc rotation mode (if used clinically) | 1 MU, 1° |
MU linearity (≥5 MU to highest MU used clinically) | ±2% | |
Accelerator output | ±1.5% | |
Coincidence of radiation and mechanical isocenter | ±1.0 mm maximum 3‐D displacement from center of target object | |
Verification of small‐field beam data — relative output factors for cones and/or MLC | ±2% from baseline for >1.0 cm apertures, ±5% from baseline for ≤1.0 cm apertures | |
E2E localization assessment “hidden target test” using SRS frame and/or IGRT system | 1.0 mm | |
E2E dosimetric evaluation using SRS frame and/or IGRT system | ±5% measured vs. calculated |
Tolerances are absolute accuracy, not variation from baseline, unless otherwise stated.