Table 1.

Minimum SRS‐SBRT relevant equipment QA and tolerances for C‐arm linac systems

Frequency Test Tolerance
Daily Laser localization — only if using SRS techniques relying on lasers for target localization (e.g., frame‐based SRS without X‐ray IGRT) 1 mm
Collimator size indicator for clinically relevant aperture 2 mm total
Radiation isocentricity test (limited gantry and couch positions) — maximum deviation in center of target object relative to each projection's beam central axis 1.0 mm SRS, 1.5 mm SBRT
IGRT positioning/repositioning 1 mm SRS, 2 mm SBRT
Imaging subsystem interlocks Functional
Stereotactic interlocks — cone size, backup jaws Functional
Accelerator output constancy ±3%
Monthly Radiation isocentricity test — covering complete range of gantry, couch, collimator positions used clinically — maximum deviation in center of target object relative to each projection's beam central axis
*Note: If both MLC and fixed conical collimators are used, both must be evaluated at least monthly
1.0 mm SRS, 1.5 mm SBRT
Treatment couch position indicators: relative over the maximum clinical range 1 mm/0.5°
Output constancy at relevant dose rates 2%
Annually SRS arc rotation mode (if used clinically) 1 MU, 1°
MU linearity (≥5 MU to highest MU used clinically) ±2%
Accelerator output ±1.5%
Coincidence of radiation and mechanical isocenter ±1.0 mm maximum 3‐D displacement from center of target object
Verification of small‐field beam data — relative output factors for cones and/or MLC ±2% from baseline for >1.0 cm apertures, ±5% from baseline for ≤1.0 cm apertures
E2E localization assessment “hidden target test” using SRS frame and/or IGRT system 1.0 mm
E2E dosimetric evaluation using SRS frame and/or IGRT system ±5% measured vs. calculated

Tolerances are absolute accuracy, not variation from baseline, unless otherwise stated.