Skip to main content
. 2017 Aug 8;18(5):10–21. doi: 10.1002/acm2.12146

Table 2.

Minimum equipment QA and tolerances for robotic linac systems

Frequency Test Tolerance
Daily*
*On days of clinical use
Head laser alignment check 1.0 mm
Safety interlocks Functional
Automatic QA (AQA) test*
*If the clinic has both fixed cones and Iris TM collimator, the AQA test should alternate between fixed cones and Iris TM , with each system tested at least weekly
Total targeting ≤1.0 mm from baseline, not exceeding manufacturer's specification
Accelerator output constancy ±3%
Monthly Energy constancy ±2%
Beam symmetry, relative ±3% for 40 mm field, ±4% for 60 mm field
Accelerator output constancy ±2%
Imager alignment 1 mm or center pixels ±2 pixels
IrisTM field size spot check (3 or more field sizes ≥10 mm) 0.5 mm
Picket fence for MLC (if applicable) Visual check
Quarterly E2E localization assessment
(Each tracking mode used clinically)
1.0 mm static target, 1.5 mm motion tracking
Annually Emergency Power Off (EPO) button, safety interlocks Functional
Accelerator output ±2.0%
MU linearity (>10 MU to highest MU used clinically) ±2%
Path verification ≤0.5 mm maximum per node, ≤0.3 mm average
Imager kVp accuracy, mA station exposure linearity, isopost alignment with center pixel ±10%, ±20%, and 1 mm, respectively
Beam laser and radiation beam alignment for cone, IrisTM, and MLC 0.5 mm from baseline
AQA baseline Recheck AQA baseline
Beam data verification — Relative output factors for cones, IrisTM, and/or MLC covering the range used clinically ±2% from baseline for >1.0 cm apertures, ±5% from baseline for ≤1.0 cm apertures