Minimum SBRT relevant equipment QA and tolerances for helical tomotherapy systems

Frequency Test Tolerance
Daily Red laser initialization (congruence with green laser) 1 mm
Image/laser coordinate coincidence 1 mm
Image registration/alignment 1 mm
Accelerator output constancy (rotational or static) ±3%
Monthly Transverse beam profile 1% average difference in field core
Longitudinal beam profile (each slice width) 1% of slice width FWHM
Output constancy and rotational output variation ±2%
Beam quality constancy ±1% PDD10 or TMR20 10
Red and green laser alignment 1 mm
Couch positioning accuracy 1 mm
CT dimensional accuracy 1 mm
Annually Couch speed uniformity ±2% dose nonuniformity
Couch translation per gantry rotation 1 mm per 5 cm
Accelerator output ±1.5%
Beam quality (each slice width) ±1% PDD10 or TMR20 10
Verification of small‐field beam data ±2% from baseline for >1.0 cm apertures, ±5% from baseline for ≤1.0 cm apertures
CT imaging — treatment — laser coordinate coincidence 1.0 mm
E2E localization assessment “hidden target test” 1.0 mm
E2E dosimetric evaluation ±5% measured vs. calculated

SRS is not included in the scope of this document for helical tomotherapy.

Tolerances are absolute accuracy, not variation from baseline, unless otherwise stated.