. 2016 Mar 8;17(2):145–152. doi: 10.1120/jacmp.v17i2.5984

Table 3.

Summary of the dosimetry quality audits of linear accelerators installed at nine different sites (Sn) from 1989 to 2015

Site
(Sn)
Audits Period Beam Energy (MV) No. of Audits
Mean(±)
SD
Minimum(±)
Maximum(±)
S14 2004‐2015 6 5 3.24 2.16 0.8 6.5
15 5 3.45 1.49 0.95 4.8
S15 2008‐2014 6 4 0.62 0.64 0.2 1.57
15 4 0.71 0.65 0.16 1.5
S16 1997‐2015 6 7 1.50 0.67 0.7 2.5
15 7 1.76 1.02 0.28 3.2
S17 1989‐2013 6 7 0.79 0.5 0.11 1.32
10/15 3 1.6 1.05 0.48 2.56
S18 2002‐2010 6 5 2.01 0.99 0.82 3.4
10/15 5 1.98 0.91 0.41 2.7
S19 2007‐2015 6 7 1.08 0.75 0.2 2.2
18 4 1.12 0.45 0.73 1.6
S20 1995‐2014 6 4 0.84 0.62 0.1 1.6
15 4 0.98 0.79 0.1 2.01
S21 2006‐2010 6 3 0.74 0.37 0.44 1.15
15 3 1.02 0.83 0.3 1.93
S22 2012‐2015 6 2 0.8 0.57 0.4 1.2