Table 3.
Summary of the dosimetry quality audits of linear accelerators installed at nine different sites () from 1989 to 2015
|
Site
|
Audits Period | Beam Energy (MV) | No. of Audits |
|
SD |
|
|
||||
|---|---|---|---|---|---|---|---|---|---|---|---|
| S14 | 2004‐2015 | 6 | 5 | 3.24 | 2.16 | 0.8 | 6.5 | ||||
| 15 | 5 | 3.45 | 1.49 | 0.95 | 4.8 | ||||||
| S15 | 2008‐2014 | 6 | 4 | 0.62 | 0.64 | 0.2 | 1.57 | ||||
| 15 | 4 | 0.71 | 0.65 | 0.16 | 1.5 | ||||||
| S16 | 1997‐2015 | 6 | 7 | 1.50 | 0.67 | 0.7 | 2.5 | ||||
| 15 | 7 | 1.76 | 1.02 | 0.28 | 3.2 | ||||||
| S17 | 1989‐2013 | 6 | 7 | 0.79 | 0.5 | 0.11 | 1.32 | ||||
| 10/15 | 3 | 1.6 | 1.05 | 0.48 | 2.56 | ||||||
| S18 | 2002‐2010 | 6 | 5 | 2.01 | 0.99 | 0.82 | 3.4 | ||||
| 10/15 | 5 | 1.98 | 0.91 | 0.41 | 2.7 | ||||||
| S19 | 2007‐2015 | 6 | 7 | 1.08 | 0.75 | 0.2 | 2.2 | ||||
| 18 | 4 | 1.12 | 0.45 | 0.73 | 1.6 | ||||||
| S20 | 1995‐2014 | 6 | 4 | 0.84 | 0.62 | 0.1 | 1.6 | ||||
| 15 | 4 | 0.98 | 0.79 | 0.1 | 2.01 | ||||||
| S21 | 2006‐2010 | 6 | 3 | 0.74 | 0.37 | 0.44 | 1.15 | ||||
| 15 | 3 | 1.02 | 0.83 | 0.3 | 1.93 | ||||||
| S22 | 2012‐2015 | 6 | 2 | 0.8 | 0.57 | 0.4 | 1.2 |