Table 3.

Mechanical tests

Item Test Frequency Tolerance
M1 Localizing lasers Daily 2 mm
Monthly 1 mm
M2 Optical distance indicator Daily 2 mm at isocenter
Monthly 2 mm over clinical range
M3 Jaw position indicators Daily 2 mm per jaw for single field
Monthly 2 mm per jaw for clinical range of motion
M4 Light to radiation field coincidence After Service 2 mm per jaw
M5 Leaf position accuracy Monthly 1 mm
M6 Gantry/collimator angle indicators Monthly
M7 Physical graticule (port film graticule) Monthly 2 mm
M8 Cross‐hair centering Monthly 1 mm
M9 Treatment couch positions (absolute and relative) Monthly Abs: 2 mm and 1°;
Rel: 1 mm over 10 cm and 0.5° over 3°
M10 Radiation isocentricity (MLC/jaw radiation isocenter with collimator, gantry and couch rotation) Annual 2 mm diametera
M11 Electron applicator collimator settings/physical inspection/interlocks Annual Same as acceptance/TPS
M12 Stereotactic accessories, lockouts, cone coding Daily Functional
Annual Functional
M13 Accessory latches/interface (all slots) Annual Functional
a

For SRS‐SBRT applications, refer to the relevant AAPM Medical Physics Practice Guideline.