Table 4.
Safety tests
| Item | Test | Frequency | Tolerance |
|---|---|---|---|
| S1 | Door interlock | Daily | Functional |
| S2 | Door closing safety | After service | Functional |
| S3 | Audio/visual monitors | Daily | Functional |
| S4 | Beam‐on indicator | Daily | Functional |
| Annual | Functional (all indicators) | ||
| S5 | Anti‐collision test | Daily | Functional (single point for system function) |
| Monthly | Functional (all collision interlocks) | ||
| S6 | Safety procedures | Determined by QMP | Functional |