Skip to main content
. 2017 May 26;18(4):23–39. doi: 10.1002/acm2.12080

Table 5.

Wedge tests

Item Test Frequency Tolerance
W1 Electronic wedge check Daily Internal: functional; collimator shaped wedges: 3%
Monthly 2%
W2 Physical wedge placement accuracy Monthly 1 mm
W3 Wedge profile for 60 degree electronic wedges, all energies Annual 2% of TPS OAFs
W4 Wedge dose for collimator shaped wedges, all angles Annual 2% of TPS dose