Item Test Frequency Tolerance Applicable to clinical practice? QMP initials
Dosimetry tests
D1 Photon and electron output constancy Dailya
Monthly
Annual
3% of baseline
2% of baseline
1% of TG51
Y/N
Y/N
Y/N
D2 Photon and electron beam profile constancy Dailya
Monthly
Annual
2%
2%
2% of TPS OAFsb
Y/N
Y/N
Y/N
D3 Electron beam energy Monthly
Annual
2 mm
2 mm
Y/N
Y/N
D4 Photon beam energy Monthly
Annual
1% of PDD/TPR (relative change in value)
1% of PDD/TPR at reference depth
Y/N
Y/N
D5 Dynamic delivery control Monthly 3% of open field dose Y/N
D6 Photon MU linearity (output constancy) Annual 2% >10 MU for open field; 2% for segmented field Y/N
D7 Electron MU linearity (output constancy) Annual 2% for clinical range Y/N
D8 Photon output vs dose rate Annual 2% Y/N
D9 Photon and electron output vs gantry angle Annual 2% of IEC gantry 0° output Y/N
D10 Photon and electron OAF vs gantry angle Annual 2% of OAFs at IEC gantry 0° Y/N
D11 Arc mode (expected MU, degree) Annual 2% of MU and 2° Y/N
D12 Special procedure mode (TBI/TSET) Annual Output: same as regular beam; energy: same as regular beam; profile: same as regular beam Y/N
Mechanical tests
M1 Localizing lasers Daily
Monthly
2 mm
1 mm
Y/N
Y/N
M2 Optical distance indicator Daily
Monthly
2 mm at isocenter
2 mm over clinical range
Y/N
Y/N
M3 Jaw position indicators Daily
Monthly
2 mm per jaw for single field
2 mm per jaw for clinical range of motion
Y/N
Y/N
M4 Light to radiation field coincidence After Service 2 mm per jaw Y/N
M5 Leaf position accuracy Monthly 1 mm Y/N
M6 Gantry/collimator angle indicators Monthly Y/N
M7 Physical graticule (port film graticule) Monthly 2 mm Y/N
M8 Cross‐hair centering Monthly 1 mm Y/N
M9 Treatment couch positions (absolute and relative) Monthly Abs: 2 mm and 1°;
Rel: 1 mm over 10 cm and 0.5° over 3°
Y/N
M10 Radiation isocentricity (MLC/jaw radiation isocenter with collimator, gantry, and couch rotation) Annual 2 mm diameterc Y/N
M11 Electron applicator collimator settings/physical inspection/interlocks Annual Same as acceptance/TPS Y/N
M12 Stereotactic accessories, lockouts, cone coding Daily
Annual
Functional
Functional
Y/NY/N
M13 Accessory latches/interface (all slots) Annual Functional Y/N
Safety tests
S1 Door interlock Daily Functional Y/N
S2 Door closing safety After Service Functional Y/N
S3 Audio/visual monitors Daily Functional Y/N
S4 Beam‐on indicator Daily Functional Functional (all indicators) Y/NY/N
Annual
S5 Anticollision test Daily
Monthly
Functional (single point for system function)
Functional (all collision interlocks)
Y/N
Y/N
S6 Safety procedures Determined by QMP Functional Y/N
Wedge tests
W1 Electronic wedge check Daily
Monthly
Internal: functional; collimator shaped wedges: 3%
2%
Y/N
Y/N
W2 Physical wedge placement accuracy Monthly 1mm Y/N
W3 Wedge profile for 60 degree electronic wedges, all energies Annual 2% of TPS OAFs Y/N
W4 Wedge dose for collimator shaped wedges, all angles Annual 2% of TPS dose Y/N
Comprehensive review of machine settings
C1 Comprehensive review of machine settings Annual Same as acceptance/expected Y/N
a

Daily checks should be conducted for the energies used that day.

b

Tolerance is the same as what was achievable for TPS model comparison to measured data as at the time of commissioning.

c

For SRS‐SBRT applications, refer to the relevant AAPM Medical Physics Practice Guideline.

Signature of qualified medical physicist:

Date: