| Item | Test | Frequency | Tolerance | Applicable to clinical practice? | QMP initials |
|---|---|---|---|---|---|
| Dosimetry tests | |||||
| D1 | Photon and electron output constancy |
Dailya
Monthly Annual |
3% of baseline 2% of baseline 1% of TG51 |
Y/N Y/N Y/N |
|
| D2 | Photon and electron beam profile constancy |
Dailya
Monthly Annual |
2% 2% 2% of TPS OAFsb |
Y/N Y/N Y/N |
|
| D3 | Electron beam energy |
Monthly Annual |
2 mm 2 mm |
Y/N Y/N |
|
| D4 | Photon beam energy |
Monthly Annual |
1% of PDD/TPR (relative change in value) 1% of PDD/TPR at reference depth |
Y/N Y/N |
|
| D5 | Dynamic delivery control | Monthly | 3% of open field dose | Y/N | |
| D6 | Photon MU linearity (output constancy) | Annual | 2% >10 MU for open field; 2% for segmented field | Y/N | |
| D7 | Electron MU linearity (output constancy) | Annual | 2% for clinical range | Y/N | |
| D8 | Photon output vs dose rate | Annual | 2% | Y/N | |
| D9 | Photon and electron output vs gantry angle | Annual | 2% of IEC gantry 0° output | Y/N | |
| D10 | Photon and electron OAF vs gantry angle | Annual | 2% of OAFs at IEC gantry 0° | Y/N | |
| D11 | Arc mode (expected MU, degree) | Annual | 2% of MU and 2° | Y/N | |
| D12 | Special procedure mode (TBI/TSET) | Annual | Output: same as regular beam; energy: same as regular beam; profile: same as regular beam | Y/N | |
| Mechanical tests | |||||
| M1 | Localizing lasers |
Daily Monthly |
2 mm 1 mm |
Y/N Y/N |
|
| M2 | Optical distance indicator |
Daily Monthly |
2 mm at isocenter 2 mm over clinical range |
Y/N Y/N |
|
| M3 | Jaw position indicators |
Daily Monthly |
2 mm per jaw for single field 2 mm per jaw for clinical range of motion |
Y/N Y/N |
|
| M4 | Light to radiation field coincidence | After Service | 2 mm per jaw | Y/N | |
| M5 | Leaf position accuracy | Monthly | 1 mm | Y/N | |
| M6 | Gantry/collimator angle indicators | Monthly | 1° | Y/N | |
| M7 | Physical graticule (port film graticule) | Monthly | 2 mm | Y/N | |
| M8 | Cross‐hair centering | Monthly | 1 mm | Y/N | |
| M9 | Treatment couch positions (absolute and relative) | Monthly |
Abs: 2 mm and 1°; Rel: 1 mm over 10 cm and 0.5° over 3° |
Y/N | |
| M10 | Radiation isocentricity (MLC/jaw radiation isocenter with collimator, gantry, and couch rotation) | Annual | 2 mm diameterc | Y/N | |
| M11 | Electron applicator collimator settings/physical inspection/interlocks | Annual | Same as acceptance/TPS | Y/N | |
| M12 | Stereotactic accessories, lockouts, cone coding |
Daily Annual |
Functional Functional |
Y/NY/N | |
| M13 | Accessory latches/interface (all slots) | Annual | Functional | Y/N | |
| Safety tests | |||||
| S1 | Door interlock | Daily | Functional | Y/N | |
| S2 | Door closing safety | After Service | Functional | Y/N | |
| S3 | Audio/visual monitors | Daily | Functional | Y/N | |
| S4 | Beam‐on indicator | Daily | Functional Functional (all indicators) | Y/NY/N | |
| Annual | |||||
| S5 | Anticollision test |
Daily Monthly |
Functional (single point for system function) Functional (all collision interlocks) |
Y/N Y/N |
|
| S6 | Safety procedures | Determined by QMP | Functional | Y/N | |
| Wedge tests | |||||
| W1 | Electronic wedge check |
Daily Monthly |
Internal: functional; collimator shaped wedges: 3% 2% |
Y/N Y/N |
|
| W2 | Physical wedge placement accuracy | Monthly | 1mm | Y/N | |
| W3 | Wedge profile for 60 degree electronic wedges, all energies | Annual | 2% of TPS OAFs | Y/N | |
| W4 | Wedge dose for collimator shaped wedges, all angles | Annual | 2% of TPS dose | Y/N | |
| Comprehensive review of machine settings | |||||
| C1 | Comprehensive review of machine settings | Annual | Same as acceptance/expected | Y/N | |
Daily checks should be conducted for the energies used that day.
Tolerance is the same as what was achievable for TPS model comparison to measured data as at the time of commissioning.
For SRS‐SBRT applications, refer to the relevant AAPM Medical Physics Practice Guideline.
Signature of qualified medical physicist:
Date: