Average gamma passing rates for ArcCHECK measurements from the Single Beam Experiment

Average gamma pass rate ± 1σ, %
Uncorrected SNC corrected New Alg. corrected
3%/3 mm 98.6 ± 3.8 98.8 ± 3.4 99.7 ± 0.9
2%/2 mm 91.7 ± 17.3 90.2 ± 15.6 95.9 ± 10.1
1%/1 mm 62.8 ± 21.2 53.9 ± 17.3 71.6 ± 19.5