Skip to main content
. 2016 Mar 8;17(2):279–290. doi: 10.1120/jacmp.v17i2.6040

Table 3.

Density difference along image profiles from Fig. 3. Values are the relative difference to kV CT in percent

Profile 1 Profile 2 Profile 3
kV CBCT (TrueBeam)
0.70±1.38
0.71±2.13
0.85±3.32
kV CBCT (Versa)
4.86±3.32
3.99±6.61
8.05±2.77
MV CBCT (Artiste)
0.80±2.74
4.76±6.83
0.06±3.73
MV CT (Tomo)
3.30±4.49
3.01±4.69
3.40±7.00