. 2014 Mar 6;15(2):205–225. doi: 10.1120/jacmp.v15i2.4685

Table 1.

Summary of required beam measurements

Algorithm
Parameter Pencil Beam Monte Carlo
Central Axis Profiles Percent Depth Dose in water at 100 cm SSD for square fields 1×1 to 15×15cm2 Percent Depth Dose in water at 90 cm SSD for fields 1×1 to 15×x15cm2
CAX profiles in air for source‐to‐detector distances (SDD) of 85 to 115 cm for fields 1×1 to 15×15cm2
Off‐Axis Profiles Diagonal Radial Profile in water at 100 cm SSD at depths from 0.5 to 350 cm, for a fully open 15×15cm2 field Inline and crossline profiles in water for 90 and 100 cm SSD at depths of 1.5, 10, and 20 cm, for square and rectangular fields 1×1 to 15×15cm2
Inline and crossline profiles in air for source‐to‐detector distances (SDD) of 85, 100, and 115 cm for square and rectangular fields 1×1 to 15×15cm2
Scatter Factors Measurements in water at 100 cm SSD and a depth of 10 cm for square fields 1×1 to 15×15cm2 relative to a 10×10cm2 field Measurements in water at 90 cm SSD and a depth of 10 cm for square and rectangular fields 1×1 to 15×15cm2 relative to a 10×10cm2 field
Measurements in air for a source‐to‐detector distance (SDD) of 100 cm for square and rectangular fields 1×1 to 15×15cm2 relative to a 10×10cm2 field
MLC Leakage Measurement in water at 100 cm SSD and a depth of 10 cm with MLC completely closed relative to an open 15×15cm2 field
Absolute Dose Gy/MU in water at 100 cm SSD, at a depth of 10 cm, for a 10×10cm2 field Gy/MU in water at 90 cm SSD, at a depth of 10 cm, for a 10×10cm2 field