Table 3.

Summary of target coverage and OAR constraints required for the RPC phantoms, with the planning/delivery technique in each case

Phantom RPC Thorax‐Lung RPC Spine RPC Head & Neck IMRT
Planning / Delivery Technique 10 Noncoplanar Conformal Fields 13 Coplanar IMRT Fields 9 Coplanar IMRT Fields
Prescription / Target Coverage >95% of PTV receives 6.00 Gy >95% of PTV receives 6.00 Gy >95% of 1° PTV receives 6.60 Gy, <1% of 1° PTV receives <6.14Gy
>99% of PTV receives 5.40 Gy 100% of PTV receives 5.25 Gy >95% of 2° PTV receives 5.40 Gy, <1% of 2° PTV receives <5.02Gy
No hotpots outside PTV No Dose >6.3Gy outside PTV
OAR Constraints Maximum Spinal Cord Dose 5.00Gy <0.35cc Spinal Cord receives >3.75Gy Maximum OAR Dose <4.5Gy
<37% Total Lung Volume receives >2.00Gy Maximum Spinal Cord Dose 5.25Gy Normal Tissue Maximum 7.26Gy
<33% Heart Volume receives >6.00Gy <1.2cc Spinal Cord receives >2.63Gy
<67% Heart Volume receives >4.50Gy <15cc Heart receives >6.00Gy
Maximum Heart Dose <4.50Gy Maximum Heart Dose <8.25Gy
Esophagus Maximum Dose 6.00Gy
<5cc Esophagus receives >4.46Gy
<1000cc Total Lung receives >2.78Gy